Residual factor Rcullis,acen for acentric reflections for this
               derivative.
               The Cullis R factor was originally defined only for centric
               reflections.  It is, however, also a useful statistical
               measure for acentric reflections, which is how it is used in
               this data item.
                           sum| |Fphobs +/- Fpobs| - Fhcalc |
          Rcullis,acen = ----------------------------------------
                                   sum|Fphobs - Fpobs|
               Fpobs  = the observed structure-factor amplitude of the native
               Fphobs = the observed structure-factor amplitude of the
                          derivative
               Fhcalc = the calculated structure-factor amplitude from the
                          heavy-atom model
               sum is taken over the specified reflections
               Ref: Cullis, A. F., Muirhead, H., Perutz, M. F., Rossmann, M. G.
                    & North, A. C. T. (1961). Proc. R. Soc. London Ser. A,
                    265, 15-38.